Home Drug Testing Home Drug Testing
Home Drug Testing
Home Drug Tests

 MORE INFORMATION
  What's New
Monthly Specials
Individual Consumers
Secure Ordering
Health Care Professionals
School Drug Testing
Workplace Testing
Corporate Programs
International Inquiries



Mononucleosis Test

A Rapid Visual Test for the Qualitative Detection of Infectious Mononucleosis Heterophile Antibody.
This test is NOT AVAILABLE FOR RETAIL SALE.
FOR HEALTHCARE PROFESSIONAL PURCHASES ONLY.

INTENDED USE

The Mono Test is a rapid test for the visual, qualitative detection of Heterophile antibodies specific to Infectious Mononucleosis (IM) in human serum, plasma or whole blood. This test kit is intended as an aid in the diagnosis of IM in patients with characteristic clinical symptoms, and is intended for profes-sional laboratory use only. This information can be used by the physician and the patient for disease management.

<< back to top

SUMMARY

Infectious Mononucleosis is an acute, self-limiting disease caused by the Epstein-Barr virus (EBV). Infection with EBV in early life usually is asymptomatic. However, up to 50% of infection occurring in young adulthood and adolescence will develop clinical manifestations associated with IM. Diagnosis of IM is based on the evaluation of characteristic clini-cal symptoms and serological changes. Serological diagnosis of IM has been demonstrated by the detection of heterophile and EBV specific antibodies. The Heterophile antibody is detectable at some point during IM in most adults. It is a widely accepted practice among physicians to use the detection of het-erophile antibodies as an aid in the diagnosis of IM. The Mono Test utilizes bovine erythrocyte extract which has a higher sensitivity and specificity than extracts from other species.

<< back to top

TEST PRINCIPLE

The Mono Test has been designed to detect IM through visual interpretation of color development in the test device, which is a sandwich solid phase gold conjugate immunoassay. The test device contains a membrane strip which is pre-coated with heterophile antigens on the test band region and goat anti-mouse antibody on the control band region. The anti-human IgM antibody-colloidal gold conjugate pad is placed at the end of the membrane. A mix-ture of colloidal gold conjugate together with the sample and developer buffer will move along the membrane chromato-graphically by capillary action. When the IM heterophile anti-bodies are present in the patient sample, the mixture will migrate to the test band region and form a visible line as the antibody complexes with the heterophile antigen. When IM heterophile antibodies are absent from the sample, no visible color band will form on test line region. Therefore, the presence of a colored band on the test line region indicates a positive result. A colored band will always appear at the control region. This control band serves as a procedural indicator for the proper performance of the test and the device.

<< back to top

STORAGE AND STABILITY

The test kit is to be stored at refrigerated (2-8°C) or at room tem-perature (up to 30°C) in the sealed pouch for the duration of the shelf-life.

<< back to top

PRECAUTIONS

  • FOR IN-VITRO DIAGNOSTIC USE ONLY.
  • Do not use test kit beyond expiration date.
  • Do not mix reagents from different lots.
  • All patient samples should be treated as if capable of transmitting disease.
  • Do not use whole blood stored for more than three days.
  • Grossly hemolyzed samples should not be used.
  • Developer Buffer and Controls contain sodium azide. Sodium azide may react with lead or copper plumbing to form potentially explosive metal azides. When disposing of these solutions, always flush with copious amounts of water to prevent azide buildup.
  • Warning: Potential Biohazardous Material Each donor unit of human plasma or serum used in the preparation of the Positive and Negative Controls was tested by FDA-approved methods for the presence of anti-HIV-1/HIV-2, HBsAg and anti-HCV, and found to be negative. However, caution should be used when handling and disposing of these items at biosafety level 2, as recommended in the Center for Disease Control/National Institutes of Health Manual, Biosafety in Micobiological and BiomedicalLaboratories, 1984.

<< back to top

REAGENTS AND MATERIALS SUPPLIED

  • 25 individually wrapped test devices. Each test cassette contains one test strip with heterophile antigen coated membrane and colored antibody pad.
  • Developer Buffer (8 ml): 0.1 M Phosphate Buffered Saline, with additives and 0.1% sodium azide.
  • Mono Negative Control (0.2 ml): Normal human plasma or serum diluted in saline solution with 0.2% sodium azide.
  • Mono Positive Control (0.2 ml): IM heterophile antibody positive human plasma or serum diluted in saline solution with 0.2% sodium azide.
  • 25 disposable Transfer Pipets.
  • One Instruction Sheet.

<< back to top

SPECIMEN COLLECTION AND HANDLING

Fingerstick

1. Clean the area to be lanced with an alcohol swab.
2. Squeeze the end of the fingertip and pierce with a sterile lancet.
3. Wipe away the first drop of blood with sterile gauze or cotton.
4. Allow the second drop to flow directly into the sample well of the test device or use the pipette provided to obtain fresh blood. Add 2 drops (about 40 µl) into the sample well.

Whole Blood

1. A certified phlebotomist should collect whole blood into a purple, blue or green top collection tube (containing EDTA, citrate or heparin, respectively) by venipuncture.
2. The whole blood may be used for testing immediately or may be stored at 2-8°C up to three days.

Plasma

1. A certified phlebotomist should collect whole blood into a purple, blue or green top collection tube (containing EDTA, citrate or heparin, respectively) by venipuncture.
2. Separate the plasma by centrifugation.
3. Carefully withdraw the plasma for testing or label and store at 2-8°C for up to two weeks. Plasma may be frozen at -20°C for up to one year.

Serum

1. A certified phlebotomist should collect whole blood into a red top collection tube (containing no anticoagulants) by venipuncture.
2. Allow the blood to clot and separate serum by centrifugation.
3. Carefully withdraw the serum for testing, or label and store at 2-8°C for up to two weeks. Serum may be frozen at -20°C for up to one year.

<< back to top

TEST PROCEDURE

Assay Procedure
1. Review specimen collection instructions.
Test device, Developer Buffer, patient's samples, controls should be brought to room temperature (20 to 30°C) prior to testing. Do not open pouches until ready to perform the assay. Remove the test device from its protective pouch. (Bring the device to room temperature before opening to avoid condensation of moisture on the membrane). Label the device with patient or control identification.
2. Add specimen to sample well:
a). For fingertip blood: Allow the second drop to flow directly into the sample well of the test device or use the pipet provided to obtain fresh blood. Add 2 drops (about 40 µl) into the sample well.
b). For whole blood samples in collection tubes: add two drops (about 40 µl) into the sample well, holding the provided transfer pipet in a vertical position.
c). For plasma or serum samples: add 1 drop (about 20 µl) of serum or plasma into the sample well, holding the provided transfer pipet in a vertical position.
3. Immediately add 3 - 4 drops of Developer Buffer.
4. After the addition of the Developer Buffer wait for the pink-red colored bands to appear. Depending on the concentration of heterophile antibodies present, positive results may be observed within 3 minutes. However, to confirm a negative result, the complete reaction time of 5 minutes is required. Do not read test results after 8 minutes.

<< back to top

INTERPRETATION OF RESULTS

1. POSITIVE: Two pink-red colored bands appear. One in the control region (C) and one pink-red colored band in the test region (T). When testing with strong positive samples, the intensity of the control band may be lighter than expected. It is not recommended to compare the intensity of the lines.

2. NEGATIVE: Only one pink-red colored band appears in the control line region. No apparent faint pink or red colored band on the test line region (T).

3. INVALID: A total absence of pink colored bands in both regions is an indication of procedural error or that test reagent deterioration has occurred.

<< back to top

QUALITY CONTROL

Internal Procedural Control
1. A procedural control is included in the test. A colored band appearing in the control region (C) is considered an internal procedural control, indicating proper performance and reactive reagents.
2. A clear background in the result window is considered an internal negative control. However, when whole blood sam-ples are tested, the background may appear slightly reddish due to the low level hemolysis of some red blood cells. This is acceptable as long as it does not interfere with the reading of the test. The test is invalid if the background fails to clear and obscures the reading of the test result.
External Quality Control
1. Positive Control: Add 1 drop (about 20 µl) of Positive Control into the sample well using the (provided) transfer pipet by holding the pipet in a vertical position. Immediately add 3 to 4 drops of the Developer Buffer. A positive signal is indicated by the development of two pink to red colored bands in the test region (T) and control region (C).
2. Negative Control: Add 1 drop (about 20 µl) of Negative Control into the sample well using the (provided) transfer pipet by holding the pipet in a vertical position. Immediately add 3 to 4 drops of the Developer Buffer. A negative signal is indicated by the development of only one pink to red colored band in control region (C).

<< back to top

LIMITATIONS

1. This test kit is to be used for the qualitative detection of IgM antibodies to IM heterophile antigen. A positive result suggests the presence of IgM antibodies to heterophile antigen.
2. This test kit should be used for symptomatic individuals suspected of having IM. Diagnosis of IM should be made by confirmation with other clinical findings.
3. A negative result does not rule out the possibility of IM because the antibodies to heterophile antigen may be absent or may not be present in sufficient quantity to be detected.

<< back to top

EXPECTED RESULTS

1. During the acute phase of IM, heterophile antibodies are detectable in 80-85% of patients. Heterophile antibodies are detectable during first month of illness and decrease rapidly after week four.
2. Positive results may be persistent for months or even years.
3. Some segment of the population who contract IM do not produce measurable level of heterophile antibodies. Approximately 50% of children under 4 years old who have IM may test negative.

<< back to top

PERFORMANCE CHARACTERISTICS

A. Accuracy
The accuracy of the Mono Test was evaluated in com-parison to a commercially available qualitative color immuno-chromatographic assay for the detection of IM IgM heterophile antibodies in human serum, plasma or whole blood. Four hun-dred and six human serum, plasma and whole blood samples (288 serum, 103 plasma and 15 whole blood) were used in the comparison study. The comparison results are summarized in Table 1.

The results indicated that the Mono Test, demonstrated a Sensitivity of 98.6% (146/148), Specificity of 98.8% (255/258), and a total agreement of 98.8% (401/406).

B. Precision/ Site Study
The precision of the Mono Test has been evaluated at ABI and at two other sites, including a physician's office and an independent clinical laboratory. Out of 27 positive samples with different levels, all (27) results were positive. Out of 33 negative samples, all (33) results were negative. The results obtained from all site studies demonstrated 100% agreement with the expected result. In addition, one posi-tive and one negative control were tested each day. Results showed 100% agreement for control specimens.

C. Sensitivity:
Since there is no sensitivity standard established for IM heterophile anti-bodies, the following dilution (test sensitivity) studies were per-formed for comparison purposes. Two mono positive human sera purchased from suppliers were used for the serial dilution in a mono negative human serum. The results of the test sensitivity study are summarized in Tables 2a and 2b.

<< back to top

 

 

Home    Products   Home Drug Testing   Order   About Us   Contact Us
Copyright © 2001 - 2009, Global Medical Systems, Inc. All Rights Reserved.